Giá bán : thỏa thuận, vui lòng liên hệ: 0976 275 983/ 0941 88 99 83, email: midtechvn@gmail.com - Mr. Hoàng Anh Quý!
Chi tiết:
 Radioactive contamination control of water and bottom sediments at depths up to 500 meters with GPS-referencing.
 
    
        
- Determination of spatial position of detection device during measurement
 - Cable reel with a current feedthrough
 - Display of measurement results on index maps of controlled radionuclides concentration or gamma radiation dose rate distribution
 - Expert mode for instrument spectrum analysis with automatic identification of sample radionuclide content
 - Expert "GARM" software for further data processing and analysis, and radiological mapping
 
| 
             AT6104DM 
             | 
            
             AT6104DM1 
             | 
        |
| Scintillation detector | 
             NaI(Tl) Ø63x63 mm 
             | 
            
             NaI(Tl) Ø63x160 mm 
             | 
        
| Energy range | 
             70 keV – 3 MeV 
             | 
        |
| Identified radionuclides | 
             137Cs, 134Cs, 131I, 40K, 226Ra, 232Th 
             | 
        |
| 
             Extended library (add 60Co, 24Na, 54Mn, etc.) available on request 
             | 
        ||
| 
             Measurement range of specific activity in water (4π geometry) 
             | 
            
             3 – 1·106 Bq/kg [134Cs, 137Cs, 131I] 
            250 – 2·104 Bq/kg [40K]  | 
            
             1 – 1·106 Bq/kg [134Cs, 137Cs, 131I] 
            100 – 2·104 Bq/kg [40K]  | 
        
| 
             Extended library (add 60Co, 54Mn, etc.) available on request 
             | 
        ||
| 
             Measurement range of specific activity in bottom sediments (2π geometry) 
             | 
            
             50 – 1·106 Bq/kg [134Cs,137Cs] 
            250 – 2·104 Bq/kg [40K]  | 
            
             – 
             | 
        
| Typical resolution at 662 keV (137Cs) | 
             7.5% 
             | 
            
             8.5% 
             | 
        
| Measurement range of ambient dose equivalent rate | 
             0.03 – 130 µSv/h 
             | 
            
             0.03 – 50 µSv/h 
             | 
        
| Limits of tolerable intrinsic relative error | 
             ±20% (for specific activity and dose rate measurement) 
             | 
        |
| Typical sensitivity to 137Cs gamma radiation | 
             2350 cps/(µSv·h-1) 
             | 
            
             5100 cps/(µSv·h-1) 
             | 
        
| Protection class of the detection device | 
             IР68 (Withstands static hydraulic pressure up to 5 MPa for not less than 24 h) 
             | 
        |
| Overall dimensions and weight of detection device | 
             Ø130x510 mm, 4.5 kg 
             | 
            
             Ø130x633 mm, 6.5 kg 
             | 
        